Documentation is available for the pump at C:\LabVIEW Modules\Drivers\WPI Aladdin-1000 Syringe Pump
| RS-232C Specifications | |
|---|---|
| Baud rate | 19200, 9600, 2400, 1200, or 300 Baud |
| Start bits | 1 bit |
| Stop bits | 1 bit |
| Parity | None |
| Data length | 8 bit |
| Flow control | Hardware |
Command Syntax:
- Communication command is terminated with .
- Communication return starts with and is terminated with .
Multiple pumps can be daisy chained together and controlled via a single IOC. Each pump has an address which can be set using the ID:SP record. By default (and in a single pump configuration) this is 00.
The device has it's own logic for setting the units of volume to be pumped VOLUME:UNITS based on the diameter of the syringe installed in the pump DIAMETER. In testing it was found that: Diameter <= 14.00mm - uL, Diameter > 14.00mm = mL. A small warning about this has been placed on the OPI to inform users.
The units for the rate can be selected but will only be set when setting the RATE:SP. You can resend the same set point to set newly selected rate units.