Skip to content

WPI Aladdin Syringe Pump

Liam edited this page May 20, 2019 · 19 revisions

Wiki > The Backend System > Specific Device IOC > Gas and liquid handling systems > WPI Aladdin Syringe Pump

Documentation

Documentation is available for the pump at C:\LabVIEW Modules\Drivers\WPI Aladdin-1000 Syringe Pump

Connection Details

RS-232C Specifications
Baud rate 19200, 9600, 2400, 1200, or 300 Baud
Start bits 1 bit
Stop bits 1 bit
Parity None
Data length 8 bit
Flow control Hardware

Command Syntax:

  • Communication command is terminated with .
  • Communication return starts with and is terminated with .

Specifications

Networked Pumps

Multiple pumps can be daisy chained together and controlled via a single IOC. Each pump has an address which can be set using the ID:SP record. By default (and in a single pump configuration) this is 00.

Volume Units

The device has it's own logic for setting the units of volume to be pumped VOLUME:UNITS based on the diameter of the syringe installed in the pump DIAMETER. In testing it was found that: Diameter <= 14.00mm - uL, Diameter > 14.00mm = mL. A small warning about this has been placed on the OPI to inform users.

Rate Units

The units for the rate can be selected but will only be set when setting the RATE:SP. You can resend the same set point to set newly selected rate units.

Clone this wiki locally