Documentation is available for the pump at \\ISIS\Shares\ISIS_Experimental_Controls\Manuals\knauer 1050 HPLC
The 1050 was designed to provide exceptionally precise and reliable solvent delivery for a wide range of HPLC applications. The pump can be fitted with either a 10 or 50 ml/min pump heads that will be RFID recognised for auto set-up. Depending on the head used there are maximum flow rates, pressures and accuracies:
| 10 ml/min | 50 ml/min | |
|---|---|---|
| Flow rate range | 0.001 to 9.000 | 0.01 to 50.00 ml/min |
| Max. delivery pressure | 400 bar (40 MPa, 5 800 psi) < 10 ml/min | 200 bar (20 MPa, 2 900 psi) < 25 ml/min |
| 150 bar (15 MPa, 2 200 psi) < 50 ml/min | ||
| Flow rate accuracy | deviation ≤ 1 % | deviation ≤ 1 % |
| Flow rate precision | RSD ≤ 0.1 % | RSD ≤ 0.3 % |