HTS Magnet
Wiki > The Backend System > Specific Device IOC > Power Supplies > HTS Magnet
Consists of an electromagnet powered by two KEPCO PSUs in master/slave configuration with monitoring by a Smartmonitor.
The HTS smart monitor is a device that shows the current status of a KEPCO for use with the HTS magnet system. It is capable of showing safe operating limits, hard limits and the current voltages and temperatures of the connected KEPCO. Currently it sits on a static IP which is set by the device, and on POLREF it is connected to the private network.
The device can be configured with its webserver which should be at <static ip>:8080; this can be used to configure a different static IP address or to enable DHCP.
The status command response on the smart monitor does not match what we have written in the "manual" (which is just an email chain) and adds a timestamp before the last status character - #7397 addresses this issue.
The Smartmonitor is a READ ONLY device which monitors (& controls?) the PSUs and magnet. Its only connection to the control system is via Ethernet on the private instrument network. It has a fixed(?) IP address of 192.168.0.11, which is not on the same subnet as the private network and so may need additional configuration of the NDX to enable communication.
The master KEPCO PSU (the uppermost in the rack) is connected to a MOXA NPort socket via a bespoke cable (NOT standard MOXA cable) with RJ45 (8P8C) on one end and RJ11 (6P6C) on the other. This flat grey cable is supplied with the device and is labelled accordingly (believed to be the only one and to date, no working spares exist). The socket on the KEPCO is labelled "Trigger" and is in the bottom left of the rear panel. The slave PSU is connected to the master, with both connected to the Smartmonitor.
This system is also used with an SMC100 single axis stage to position the sample inside the magnet itself.