@@ -11,18 +11,18 @@ The following faults can be seen when operating the magnet fully from the front
1111
* The switch heater occasionally reports that it's ON when it's actually OFF
1212
* The power supply reports a voltage of ~9000V which is incorrect (a sensible voltage for this power supply would be around ~8V while ramping)
1313

14-
## SCPI Prorocol
14+
## SCPI Protocol
1515

1616
The IPS IOC now supports the SCPI protocol, which is more feature rich than Legacy mode.
17-
Effort was made to ensure that the top level EPICS insterface was was changed as little as possible.
17+
Effort was made to ensure that the top level EPICS interface was was changed as little as possible.
1818
It was particularly important that the SNL state-machine logic was not altered.
1919
This all required some careful shoe-horning of the new interface to provide PV compatibility with the legacy mode.
20-
There is now siginificantly more diagnostic information available, along with support for daughter-boards, such as He and N leve meters, pressure measurement, etc., all of which have necessitated additions to the user interface in the IBEX client.
20+
There is now significantly more diagnostic information available, along with support for daughter-boards, such as He and N leve meters, pressure measurement, etc., all of which have necessitated additions to the user interface in the IBEX client.
2121
The IOC can be configured to run with either legacy or SCPI protocols via a STREAMPROTOCOL macro ("SCPI" | "LEGACY"). The IOC publishes a new PV $(P)PROTOCOL, which reflects the configuration mode, allowing the user interface to hide or show attributes and controls relevant to SCPI or LEGACY modes.
2222

2323
As already mentioned, SCPI mode provides additional status reporting, much of which is based on the return string from the "READ:SYS:ALRM" command, which is poorly documented in the supplier's documentation. The status strings are assumed to all conform to the "Directory of Alarms" section (17.3) of the Operator's Manual (Issue 20, July 2018).
2424
Whilst many of the system alarms that we could test, mostly conformed, some differences were noticed, along with additional, undocumented status, such as "Magnet Safety".
25-
It has not been possible to test all alarm scenarios with the IPS unit and as such unable to fully assertain that all the expected message strings are correct - they may need to be adjusted later on, if/when they arise.
25+
It has not been possible to test all alarm scenarios with the IPS unit and as such unable to fully ascertain that all the expected message strings are correct - they may need to be adjusted later on, if/when they arise.
2626

2727
The support module exports an aSub record subroutine to facilitate handling of the responses to READ:SYS:ALRM, which is not feasible with a StreamDevice protocol handler.
2828